Arrhythmias & electrophysiology

Safety and performance of conduction system pacing: Real-world experience from a product surveillance registry

Heart RhythmPublished Observational registry analysis

Authors' conclusion

In a multicenter cohort of LBBAP and HBP patients treated with the catheter-delivered 3830 lead, lead-related complication rates were low and electrical parameters were stable through 30 months.

Study context

Population
Patients undergoing pacemaker implantation for conduction system pacing (CSP) with Medtronic 3830 lead, Medtronic product surveillance registry (PSR), 77 centres, Jan 2019–Dec 2023
Participants
2,342 patients · 77 centres
Exposure
Left bundle branch area pacing (LBBAP) vs His-bundle pacing (HBP) with the Medtronic 3830 lead
  • LBBAPn = 1,502
  • HBPn = 840reference
Outcome
Lead-related complications and pacing capture threshold
Methodological detailsmulticentre, product surveillance registry
  • Multicentre
  • Product surveillance registry

Main finding

As reported in the abstract

Lead complication rate · 36 months postimplant
LBBAP had lower pacing thresholds than HBP through 30 months
P < 0.001

Observational data: associations, not effects.

Further findings

R-wave sensing and impedance · through 30 months

LBBAP had higher R-wave sensing and higher impedance than HBP
P < 0.001

All-cause mortality · 36 months

No difference in all-cause mortality

Original publication

Safety and performance of conduction system pacing: Real-world experience from a product surveillance registry. Heart Rhythm. Published 3 July 2024. doi:10.1016/j.hrthm.2024.06.061

Summarised from the published abstract; every number is checked against it