Arrhythmias & electrophysiology

Safety of magnetic resonance imaging in patients with surgically implanted permanent epicardial leads

Heart RhythmPublished Prospective cohort study

Authors' conclusion

MRI of CIEDs with epicardially implanted leads does not represent a greater risk than transvenous CIEDs when performed with a multidisciplinary collaborative protocol centered on patient safety.

Study context

Population
Patients with surgically implanted epicardial cardiac implantable electronic device (CIEDs) undergoing magnetic resonance imaging (MRI) under a cardiology-radiology collaborative protocol, 2008–202152 MRIs in epicardial group
Exposure
Epicardial CIED vs matched non-MRI-conditional transvenous CIED undergoing MRI
  • Epicardial CIEDn = 29
Outcome
Safety and adverse effects of MRI
Methodological details2 clinical centres, matched comparison group
  • 2 clinical centres
  • Matched comparison group

Main finding

As reported in the abstract

Significant adverse events

Epicardial CIED

0

events

Transvenous CIED

0

events

Observational data: associations, not effects.

Further findings

Battery life, pacing, sensing thresholds, lead impedance, cardiac biomarkers

Not significantly changed, except 1 patient with transient decrease in atrial lead sensing

Safety

Events

OutcomeEpicardial CIEDTransvenous CIED
Significant adverse events00

Battery life, pacing, sensing thresholds, lead impedance, cardiac biomarkers

Not significantly changed, except transient decrease in atrial lead sensing function (patients)
1

Original publication

Safety of magnetic resonance imaging in patients with surgically implanted permanent epicardial leads. Heart Rhythm. Published 17 April 2023. doi:10.1016/j.hrthm.2023.04.003

Composed by AI from the published abstract and checked against it. How Briefs are made →