Heart failure & cardiomyopathies

Twelve-Month Follow-Up Results from the SIRONA 2 Clinical Trial

ESC Heart FailPublished Prespecified secondary endpoint analysis of the SIRONA 2 single-arm trial

Authors' conclusion

Long-term follow-up of the SIRONA 2 trial supports the safety and accuracy of the Cordella PA Sensor System in enabling comprehensive HF management in NYHA class III HF patients.

Study context

Population
New York Heart Association (NYHA) class III heart failure (HF) with a prior HF hospitalization and/or increase of NT-proBNP within 12 months of enrolment
Participants
70 participants
Primary endpoint
Prespecified secondary endpoints: safety and accuracy of the PA sensor (vs RHC)
Follow-up
12 months
Methodological detailsprospective, multicentre, open-label
  • Prospective
  • Multicentre
  • Open-label

Primary result

As reported in the abstract

Average difference for mean PAP, sensor vs right heart catheterization (RHC) (mean ± SD)

2.9 ± 7.3 mmHg

Freedom from device/system-related complications

98.4%

Pressure sensor failures

0

Secondary outcomes

HF hospitalizations and mortality · 12 months

Event rate per patient year
0.33

NYHA symptoms · 12 months

Significantly improved
P < 0.0001

Functional capacity (6 min walk test) · 12 months

Significantly improved
P = 0.02

Adherence to daily transmissions of PAP and vital signs · 12 months

Adherence
95%

Safety

Device/system-related complications · 12 months

Freedom from device/system-related complications
98.4%
Pressure sensor failures
0

Original publication

Twelve-Month Follow-Up Results from the SIRONA 2 Clinical Trial. ESC Heart Fail. Published 25 January 2024. doi:10.1002/ehf2.14657

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.