Ventricular Fibrillation During Pulsed Field Ablation With Dual-Energy Focal Catheters in Patients With Cardiac Implantable Electronic Devices
Authors' conclusion
Unexpected VF during PFA in patients with CIEDs occurred in 3.7% of procedures. These results complement prior adverse-event reports and support systematic pre-, intra-, and post-procedure CIED management, with external defibrillation immediately available, whenever PFA or dual-energy ablation is performed in patients with cardiac devices.
Study context
- Population
- Patients with cardiac implantable electronic devices (CIEDs) undergoing pulsed field ablation (PFA)consecutive procedures using PFA energy · 83 (atrial ablations) · 24 (ventricular ablations)
- Participants
- 107 consecutive procedures
Methodological detailsconsecutive PFA procedures
- Consecutive PFA procedures
Further findings
Device and intra-procedural findings in ventricular fibrillation (VF) cases
- VF occurred at ablation sites both near and remote from CIED lead tips
- Device electrograms demonstrated prominent ablation artifact with abnormal triggered activity and mode switch behavior
- Intra-procedural recordings also demonstrated local capture and R on T mechanisms
Safety
Unexpected ventricular fibrillation (VF) during PFA · procedural
- Unexpected VF, patients
- 4
- Overall incidence (95% CI 1.0% - 9.3%)
- 3.7%
- During left atrial PFA for pulmonary vein isolation
- 1
- During left ventricular PFA for ventricular tachycardia ablations
- 3
Original publication
Ventricular Fibrillation During Pulsed Field Ablation With Dual-Energy Focal Catheters in Patients With Cardiac Implantable Electronic Devices. Heart Rhythm. Published 6 November 2026. doi:10. 1016/ j. hrthm. 2026. 11. 004