Arrhythmias & electrophysiology

Visualization of Pulsed Field Ablation (PFA) During PVI With the Second-Generation Pentaspline Catheter: NAVIGATE-PF Phase 1 Results

JACC Clin ElectrophysiolPublished First-in-human feasibility study

Authors' conclusion

The second-generation pentaspline PFA catheter, with dynamic shape visualization and preview of anticipated electric field, resulted in alignment with postablation voltage mapping.

Study context

Population
AF patients undergoing PVI with the second-generation pentaspline PFA catheter15 in acute tags analysis
Participants
30 patients
Outcome
Estimated shape and position of acute tags relative to low-voltage borders on high-density postablation maps
Methodological detailsmulticentre
  • Multicentre

Main finding

As reported in the abstract

Post-procedural

Distance between tag and low-voltage border

−0.58 mm

Q1-Q3: −2.9 to 2.17 mm

Further findings

Successful treatment with the second-generation PFA catheter

All 30 patients successfully treated

Original publication

Visualization of PFA During PVI With the Second-Generation Pentaspline Catheter: NAVIGATE-PF Phase 1 Results. JACC Clin Electrophysiol. Published 30 December 2025. doi:10.1016/j.jacep.2025.11.016