Arrhythmias & electrophysiology

Accuracy and variability of cardiologist interpretation of single lead electrocardiograms for atrial fibrillation: The VITAL-AF trial

Am Heart JPublished Diagnostic accuracy study

Authors' conclusion

Cardiologist interpretation of point-of-care handheld 1L ECGs has modest diagnostic sensitivity and specificity with substantial variability for AF classification despite high confidence.

Study context

Population
Patients aged ≥65 years in the VITAL-AF trial with point-of-care handheld 1L ECG1L ECG tracings
Participants
200 patients
Comparison
Cardiologist interpretation of 1L ECG vs same-day 12L ECG
Outcome
Sensitivity, specificity, variability and confidence of cardiologist interpretation for AF
Methodological detailscardiologists blinded to algorithm and 12L ECG interpretation, stratified random sampling
  • Cardiologists blinded to algorithm and 12L ECG interpretation
  • Stratified random sampling

Performance

As reported in the abstract

Cardiologist interpretation of 1L ECGs for AF (vs same-day 12L ECG)

Average sensitivity for atrial fibrillation (AF)

77.4%

range 50%–90.6%, SD=11.4%

Average specificity

73.0%

range 41.3%–94.6%, SD = 15.4%

Mean variability

30.8%

range 0%–76.2%, SD = 23.2%

Average reviewer confidence (out of 5)

3.6

range 2.5–4.2, SD = 0.6

Further findings

Factors associated with sensitivity, specificity, variability and confidence

Associated factors included age, BMI and presence of artifact
  • Multivariable regression

Original publication

Accuracy and variability of cardiologist interpretation of single lead electrocardiograms for atrial fibrillation: The VITAL-AF trial. Am Heart J. Published 13 July 2023. doi:10.1016/j.ahj.2023.07.003