Valvular & structural heart disease

30-Day and 1-Year Outcomes of Navitor Transcatheter Aortic Valve in Low- or Intermediate-Risk Patients

JACC Cardiovasc IntervPublished Prospective single-arm study

Authors' conclusion

The Navitor valve demonstrated favorable safety and performance outcomes at 12 months

Study context

Population
Severe aortic stenosis (AS) at low or intermediate surgical risk undergoing Navitor TAVR, 36 sites in Europe, Australia and Israel
Participants
434 participants
Exposure
Low risk vs Intermediate risk
  • Low riskn = 203
  • Intermediate riskn = 231
Outcome
Effectiveness: moderate or greater paravalvular leak (PVL); safety: all-cause mortality or fatal stroke or stroke with disability (each vs a performance goal)
Follow-up
30 days (PVL); 12 months (safety)
Methodological detailsmulticentre, performance goal, Clinical events adjudicated by an independent committee; imaging by a core laboratory
  • Multicentre
  • Performance goal
  • Clinical events adjudicated by an independent committee; imaging by a core laboratory

Main finding

As reported in the abstract

Moderate or greater paravalvular leak (PVL) at 30 days (performance goal (PG) 6.6%, P < 0.0001)

0%

All-cause mortality or fatal stroke or stroke with disability at 12 months, first 262 patients (PG 11.3%, P < 0.0001)

2.3%

Observational data: associations, not effects.

Further findings

Technical success

Technical success
97.0%
Procedural mortality
0

Hemodynamic performance · through 12 months

Mean gradient
8.0 mm Hg
Effective orifice area
1.8 cm²

Original publication

30-Day and 1-Year Outcomes of Navitor Transcatheter Aortic Valve in Low- or Intermediate-Risk Patients. JACC Cardiovasc Interv. Published 29 August 2025. doi:10.1016/j.jcin.2025.08.021

Composed by AI from the published abstract and checked against it. How Briefs are made →