Valvular & structural heart disease

ACURATE neo2 versus SAPIEN 3 Ultra (Ultra) for transcatheter aortic valve implantation

EuroInterventionPublished Observational cohort study

Main finding · Device successResult — go to the resultOdds ratio 1.961 95% CI 1.269–3.031

Authors' conclusion

Short-term outcomes after TAVI using the Neo2 or Ultra THV were excellent and, overall, comparable. However, transvalvular gradients were lower with the Neo2, which translated into higher rates of device success. Rates of mild PVL were significantly lower with the Ultra THV.

Study context

Population
Patients undergoing transcatheter aortic valve implantation (TAVI)
Participants
1,356 patients472 matched pairs
Exposure
ACURATE neo2 vs SAPIEN 3 Ultra for TAVI
  • Neo2 (Neo2)n = 608
  • Ultran = 748reference
Outcome
Device success (latest VARC definitions)
Methodological details4 centres, propensity-score matched, inverse probability treatment weighting
  • 4 centres
  • Propensity-score matched
  • Inverse probability treatment weighting

Main finding

Device success

Odds ratio

1.961

95% CI 1.269–3.031

P = 0.002

By group

Neo2vs Ultra
OR 1.961 (95% CI 1.269–3.031); P = 0.002

As reported in the abstract

Observational data: associations, not effects.

Further findings

Moderate to severe PVL · after PSM

P = 0.725

Elevated transvalvular gradients (≥20 mmHg) · after PSM

P < 0.001

Mild PVL

P < 0.001

Safety

Clinical events · 30 days

Comparable between the 2 groups

Original publication

ACURATE neo2 versus SAPIEN 3 Ultra for transcatheter aortic valve implantation. EuroIntervention. Published 23 January 2023. doi:10.4244/eij-d-22-00164