Arrhythmias & electrophysiology

Acute results of the ASCEND EV study: A new modality of parasternal extravascular implantable cardioverter-defibrillator therapy including antitachycardia pacing

Heart RhythmPublished Pilot study

Authors' conclusion

These results show that the AtaCor parasternal EV-ICD system is feasible and safe, with high defibrillation efficacy in induced ventricular arrhythmias and a low occurrence of major complications within 30 days.

Study context

Population
Patients with a class I or IIa indication for ICD therapy
Participants
19 patients
Outcome
Acute defibrillation success and freedom from major complications
Follow-up
30 days

Main finding

As reported in the abstract

Acute defibrillation success

Defibrillation testing successful (17 of 18 tested)

94%

Lead successfully deployed (19 of 19 attempted implants)

100%

Further findings

OutcomeValues
Successful lead deploymentParasternal EV-ICD 100%
Intraprocedural pacing capture (dedicated bipolar pacing vector)Parasternal EV-ICD 100%
  • Successful lead deployment: 19 of 19 attempted implant procedures
  • Intraprocedural pacing capture (dedicated bipolar pacing vector): 17 of 17 chronically implanted subjects

Intraprocedural pacing capture (dedicated bipolar pacing vector)

Pacing well tolerated at predischarge visit

Safety

Freedom from major complications · 30 days

Freedom from major complications
95% (95% CI 74% to >99%)

Original publication

Acute results of the ASCEND EV study: A new modality of parasternal extravascular implantable cardioverter-defibrillator therapy including antitachycardia pacing. Heart Rhythm. Published 15 November 2025. doi:10.1016/j.hrthm.2025.10.035

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.