Arrhythmias & electrophysiology

Atrioventricular Synchrony Delivered by a Dual-Chamber Leadless Pacemaker System

CirculationPublished Prospective single-arm clinical trial

Authors' conclusion

This dual-chamber leadless pacemaker system demonstrated atrioventricular synchrony in 98% of evaluable beats at 3 months after implantation. Atrioventricular synchrony was maintained across postures/activities and remained robust for heart rates >100 bpm.

Study context

Population
Patients with a conventional dual-chamber pacing indication, enrolled February 2022 to March 2023384 evaluable
Participants
464 patients
Outcome
Percent of AV synchronous beats (300-ms PR interval limit) on 12-lead Holter across postures/activities, core-lab adjudicated
Follow-up
3 months after implantation
Methodological detailsunblinded, multicenter, international, post hoc analyses
  • Unblinded
  • Multicenter
  • International
  • Post hoc analyses

Main finding

As reported in the abstract

AV synchrony (% of beats, 300-ms PR limit) vs i2i communication success
Mean AV synchrony across all postures
98%
Atrial-to-ventricular i2i
94%
Ventricular-to-atrial i2i
94%
AV synchrony greater than both i2i values
P<0.001
AV synchrony exceeded both i2i values in % of patients
95%

Observational data: associations, not effects.

Further findings

AV synchrony by posture/activity, implantation indication, AV paced/sensed event combination and heart rate range (post hoc) · 3 months

AV synchrony achieved in >95% of evaluable beats across all categories, including heart rate >100 bpm
  • Post hoc

Original publication

Atrioventricular Synchrony Delivered by a Dual-Chamber Leadless Pacemaker System. Circulation. Published 8 July 2024. doi:10.1161/circulationaha.124.069006

Composed by AI from the published abstract and checked against it. How Briefs are made →