Arrhythmias & electrophysiology

Comparison of in-hospital outcomes and complications of leadless pacemaker and traditional transvenous pacemaker implantation

EuropacePublished Retrospective cohort study

Authors' conclusion

There is a higher likelihood of all-cause in-hospital mortality and complications following LPM implantation in comparison to TV-VVI.

Study context

Population
Adults undergoing leadless or single-chamber transvenous pacemaker implantation, US national inpatient database, 2016–2019expanded samples of admissions
Participants
35,430 adults
Exposure
Leadless pacemakers (LPM) vs TV-VVI
  • LPMn = 7,780
  • TV-VVIn = 27,650reference
Outcome
Procedural complications and in-hospital outcomes
Follow-up
In-hospital
Methodological detailsnational inpatient database, multivariable logistic regression, 1:1 propensity score matching
  • National inpatient database
  • Multivariable logistic regression
  • 1:1 propensity score matching

Main finding

As reported in the abstract

LPM vs TV-VVIAdjusted odds ratio (95% CI)P value
In-hospital mortality
1.63(1.29–2.05)
P < 0.001
Vascular complications
7.54(3.21–17.68)
P < 0.001
VTE
3.67(2.68–5.02)
P < 0.001
Cardiac complications
1.79(1.59–2.03)
P < 0.001
Device thrombus formation
5.03(2.55–9.92)
P < 0.001
Blood transfusion
1.54(1.14–2.07)
P < 0.005
Pulmonary complications (higher with TV-VVI)
0.68(0.54–0.87)
P < 0.002
Device revisions (higher with TV-VVI)
0.42(0.23–0.76)
P < 0.004

Observational data: associations, not effects.

Original publication

Comparison of in-hospital outcomes and complications of leadless pacemaker and traditional transvenous pacemaker implantation. Europace. Published 15 September 2023. doi:10.1093/europace/euad269

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.