Arrhythmias & electrophysiology

Extravascular Implantable Cardioverter-Defibrillator Sensing and Detection in a Large Global Population

JACC Clin ElectrophysiolPublished Secondary analysis of the extravascular (EV) implantable cardioverter-defibrillator (ICD) pivotal study

Authors' conclusion

In a global population, EV-ICD detected induced and spontaneous VT/VF accurately. Although discriminators withheld detection from most non-VT/VF episodes, inappropriate shocks were common.

Study context

Population
Patients discharged with an EV-ICD in the global pivotal study
Participants
299 patients
Primary endpoint
Sensing and detection performance (induced VF and spontaneous episodes in a programmed VT/VF therapy zone)
Methodological detailsglobal, adjudicated episodes
  • Global
  • Adjudicated episodes

Primary result

As reported in the abstract

VT/VF detection
Induced ventricular fibrillation (VF) detected at programmed sensitivity at implantation
all
Induced VF detected with 3× safety margin
95.9%
Sustained spontaneous ventricular tachycardia (VT)/VF episodes detected in follow-up
59

Further findings

Non-VT/VF episodes (n = 1,034)

Caused by oversensing
87.9%
Caused by SVT
12.1%
Therapy withheld
80.9%
Therapy aborted
10.6%
Therapy delivered
8.5%
Oversensing from myopotentials
61.2%
Oversensing from P-wave oversensing (PWOS)
19.9%

Predictors of oversensing

Myopotential oversensing more common with Ring-Can sensing
P < 0.0001
Myopotential oversensing correlated with low R-wave amplitude
P < 0.0001
PWOS almost exclusively with Ring1-Ring2 sensing
P = 0.0001
PWOS began with transient decrease in R-wave or increase in P-wave amplitude
P < 0.0001
New PWOS discriminator significantly reduced total inappropriate detections (software emulation)

Safety

Inappropriate shocks

Myopotential episodes
3.2%
PWOS episodes
21.8%

Original publication

Extravascular Implantable Cardioverter-Defibrillator Sensing and Detection in a Large Global Population. JACC Clin Electrophysiol. Published 15 May 2024. doi:10.1016/j.jacep.2024.02.033

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.