Arrhythmias & electrophysiology

First-in-human clinical series of a novel conformable large-lattice pulsed field ablation catheter for pulmonary vein isolation

EuropacePublished Single-arm first-in-human trial

Authors' conclusion

Pulmonary vein isolation (PVI) utilizing a conformable single-shot PFA catheter to treat PAF was efficient, safe, and effective, with durable lesions demonstrated upon remapping.

Study context

Population
Patients with paroxysmal AF undergoing pulmonary vein isolation (PVI) with an 8 Fr large-lattice conformable single-shot pulsed field ablation (PFA) catheter
Participants
85 patients
Comparison
PULSE1 vs PULSE2 vs PULSE3
  • PULSE1n = 30
  • PULSE2n = 20
  • PULSE3n = 35
Outcome
Acute PVI
Methodological details3 centres, 5 operators
  • 3 centres
  • 5 operators

Main finding

As reported in the abstract

Acute PVI achieved (% of PVs)

100%

PFA applications per PV (mean ± SD)

3.9 ± 1.4

Secondary outcomes

1-year freedom from atrial arrhythmias (Kaplan-Meier) · one year — PULSE3
100% (95% CI 80.6–100%)
PVI durability at remapping (per vein) · ∼75 days post-ablation — PULSE3
99%
1-year freedom from atrial arrhythmias (Kaplan-Meier) — Total cohort
81.8% (95% CI 70.2-89.2%)
PVI durability at remapping (per vein) — Total cohort
90%

Procedure times (overall procedure, transpired ablation, PFA catheter dwell, fluoroscopy)

Overall procedure time (min)
56.5 ± 21.6
Transpired ablation time (min)
10.0 ± 6.0
PFA catheter dwell time (min)
19.1 ± 9.3
Fluoroscopy time (min)
5.7 ± 3.9

Safety

No pre-defined primary safety events occurred

Original publication

First-in-human clinical series of a novel conformable large-lattice pulsed field ablation catheter for pulmonary vein isolation. Europace. Published 8 April 2024. doi:10.1093/europace/euae090

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.