Arrhythmias & electrophysiology

Intermuscular technique for implantation of the subcutaneous implantable defibrillator: a propensity-matched case–control study

EuropacePublished Retrospective cohort study

Authors' conclusion

Our data showed the superiority of the IM S-ICD generator positioning in reducing device-related complications and inappropriate shocks.

Study context

Population
Patients undergoing subcutaneous implantable cardioverter defibrillator (S-ICD) implantation, 2013–2021matched cohort
Participants
1,577 patients
Exposure
Intermuscular (IM) vs subcutaneous (SC) S-ICD generator positioning
  • IMn = 290
  • SCn = 290reference
Outcome
Survival from device-related complications and inappropriate shocks
Follow-up
Median 28 months
Methodological detailspropensity-score matched, consecutive patients
  • Propensity-score matched
  • Consecutive patients

Main finding

As reported in the abstract

Each group vs SCEstimate (95% CI)P value
IM
HR for device-related complications 0.41(0.17–0.99)
P = 0.041
IM
HR for composite of complications and inappropriate shocks 0.50(0.30–0.86)
P = 0.013

Observational data: associations, not effects.

Further findings

OutcomeHR (95% CI)P value
Appropriate shocks0.90 (0.50–1.61)P = 0.721

Interaction by gender, age, BMI, EF

No significant interaction between generator positioning and gender, age, BMI or EF

Original publication

Intermuscular technique for implantation of the subcutaneous implantable defibrillator: a propensity-matched case–control study. Europace. Published 16 February 2023. doi:10.1093/europace/euad028

Composed by AI from the published abstract and checked against it. How Briefs are made →