Leadless Pacemakers in the Setting of Surgical and Transcatheter Tricuspid Valve Procedures

Circ Arrhythm ElectrophysiolPublished

  • systematic review

Authors' conclusion

Across a broad spectrum of complex tricuspid anatomies, leadless pacemakers demonstrated excellent feasibility, durable electrical performance, and a low complication profile, supporting their role as a valve-sparing pacing strategy in this population. These findings support leadless pacing in patients with prior tricuspid intervention, although prospective comparative data are required to define its role relative to alternative pacing modalities.

Study context

Population
Patients undergoing surgical or transcatheter tricuspid valve interventions who received leadless pacemaker implantation
Participants
272 Thirty-four studies

Pooled result

As reported in the abstract

Procedural success

procedural success rate

99.3%

Further pooled results

Electrical performance · during follow-up

stable capture thresholds, sensing amplitudes, and impedance

Device-valve interaction

infrequent and generally manageable

Worsening tricuspid regurgitation attributable to the leadless pacemaker or its delivery system

no study demonstrated worsening tricuspid regurgitation attributable to the leadless pacemaker or its delivery system

Safety

Leadless pacemaker-related complications

Leadless pacemaker-related complications
1.1%
no device-related mortality

Original publication

Leadless Pacemakers in the Setting of Surgical and Transcatheter Tricuspid Valve Procedures. Circ Arrhythm Electrophysiol. Published 20 May 2026. doi:10.1161/circep.126.015000

Summarised from the published abstract; every number is checked against it