Arrhythmias & electrophysiologyCohort study

Micra leadless pacemaker revisions: incidence, characteristics, and outcomes from a multicentre French cohort

EuropacePublished

  • Retrospective cohort study
  • multicentre
  • 46 French centres

Authors' conclusion

LLP revisions are infrequent and predominantly managed without device removal.

Study context

Population
Patients with Micra leadless pacemakers implanted at 46 French centres, 2015–20238,994 Micra LLPs implanted
Participants
100 patients
Exposure
LLP revision with device removal vs without removal
  • Removaln = 30
  • Without removaln = 70reference
Outcome
Incidence, characteristics and outcomes of LLP revisions, with emphasis on feasibility and safety of removal

Main finding

As reported in the abstract

LLP revision incidence · median follow-up 229 (68.5-629.8) days
Revision procedures among 8994 Micra LLPs implanted
100 (1.1%)
Revisions managed without device removal
70
Revisions involving removal
30
Indication: device upgrade
55%
Indication: high pacing threshold
33%
Indication: battery depletion
3%
Indication: ventricular arrhythmias
3%
Indication: miscellaneous causes
6%

Observational data: associations, not effects.

Further findings

Time to revision

P < 0.001

Time to revision — Revisions occurring within 1 year
80%

Safety

Device removal success and complications
97%

Original publication

Micra leadless pacemaker revisions: incidence, characteristics, and outcomes from a multicentre French cohort. Europace. Published 11 November 2025. doi:10.1093/europace/euaf291

Summarised from the published abstract; every number is checked against it