Coronary disease & ACS

Predictive Accuracy of Magnetocardiography for Diagnosing Myocardial Ischemia in NSTE-ACS Patients With Residual Post-PCI Angina

JACC AdvPublished development and internal validation of a predictive model

Authors' conclusion

The OPM-MCG model provides a reliable, noninvasive, and quick tool for identifying post-PCI angina in NSTE-ACS patients, highlighting its clinical value for risk stratification, guiding personalized pharmacologic therapy, and follow-up strategies.

Study context

Population
NSTE-ACS patients who underwent PCI with pre- and post-PCI optically pumped magnetometer magnetocardiography (OPM-MCG) scansNSTE-ACS patients who underwent PCI
Participants
363 patientsEvents: 134 (36.9%)
Predictors
OPM-MCG model incorporating five Δ parameters
Predicted outcome
post-PCI angina (Seattle Angina Questionnaire angina frequency score <100 at 3 months)
Methodological detailsstratified bootstrap least absolute shrinkage and selection operator regression (1,000 iterations), 10-fold cross-validation
  • Stratified bootstrap least absolute shrinkage and selection operator regression (1,000 iterations)
  • 10-fold cross-validation

Model performance

As reported in the abstract

Area under curve
Average during bootstrapping
0.85(95% CI 0.78-0.93)
Final model with 10-fold cross-validation
0.79(95% CI 0.78-0.80)

Further findings

Characteristics of patients with post-PCI angina

Patients with post-PCI angina had higher body mass index, more frequent history of PCI, and greater use of oral nitrates

Original publication

Predictive Accuracy of Magnetocardiography for Diagnosing Myocardial Ischemia in NSTE-ACS Patients With Residual Post-PCI Angina. JACC Adv. Published 19 August 2025. doi:10.1016/j.jacadv.2025.102073

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.