Arrhythmias & electrophysiology

Pulsed Field Ablation Using Focal Contact Force–Sensing Catheters for Treatment of Atrial Fibrillation: 1-Year Outcomes of the ECLIPSE AF Study

Circ Arrhythm ElectrophysiolPublished Prospective single-arm study

Authors' conclusion

This study demonstrated that optimization of focal PFA with 3 contact force-sensing, solid-tip ablation catheters resulted in the progressive improvement of PV isolation durability at 3-month remapping and high freedom from atrial arrhythmia survival rates, providing a promising focal PFA treatment option integrated with current ablation workflows.

Study context

Population
Patients with paroxysmal or persistent AF undergoing pulmonary vein (PV) isolation80 (98%) remapped
Participants
82 patients
Outcome
Freedom from atrial arrhythmia
Follow-up
12 months
Methodological detailsmulticentre
  • Multicentre

Main finding

As reported in the abstract

1 year

Focal pulsed field ablation (PFA)

80.2%

95% CI 69.7%–87.4%

Observational data: associations, not effects.

Further findings

Acute PV isolation

PV isolation achieved in targeted veins
100% (322/322)
First-pass isolation
92.2% (297/322)

Durable isolation at 90-day remapping, optimized PFA cohorts 3, 4 and 5

Per-patient isolation, cohorts 3, 4, 5
60%, 73%, 81%
Per-PV isolation, cohorts 3, 4, 5
84%, 90%, 92%
Patients remapped
80 (98%)

Safety

Primary safety events

Primary safety events in 4 patients
4.9%, 4/82
Nonembolic stroke (tamponade secondary to catheter perforation)
1
Hemorrhagic vascular access complications
3
Fistula, diaphragmatic paralysis, MI, pericarditis, thromboembolism, PV stenosis, TIA or death
0

Original publication

Pulsed Field Ablation Using Focal Contact Force–Sensing Catheters for Treatment of Atrial Fibrillation: 1-Year Outcomes of the ECLIPSE AF Study. Circ Arrhythm Electrophysiol. Published 19 December 2024. doi:10.1161/circep.124.012794

Summarised from the published abstract; every number is checked against it