Arrhythmias & electrophysiologyResearch article

Vector estimation, resampling, and smoothing algorithm (VERSA): Omnipolar vector mapping from arbitrarily positioned electrodes

Heart RhythmPublished

  • Omnipolar vector mapping from electroanatomic mapping data

Authors' conclusion

Extending omnipolar technology to arbitrary electrode arrangements and visualizing measurements in VERSA maps can elucidate wavefront propagation in reentrant tachycardias without the need for LAT annotation. These methods could facilitate better interpretation of complex arrhythmia patterns and more effective ablation.

Study context

Population
Scar-related reentrant atrial tachycardias and ventricular tachycardia mapped with PENTARAY and OCTARAY catheters34 scar-related reentrant atrial tachycardias and 1 ventricular tachycardia · scar-related reentrant atrial tachycardias 34 · ventricular tachycardia 1
Comparison
Omnipolar estimates of conduction velocity (CV), activation direction (AD) and voltage compared with conventional analogs (local activation time (LAT)-derived equivalents)

Main finding

As reported in the abstract

Conduction velocity (CV)
  • omnipolar CV was higher than LAT-based CV

Further findings

Activation direction (AD)

differences in AD were negligible

Voltage

Omnipolar voltages were significantly higher than the corresponding maximum bipolar voltages

VERSA maps (qualitative assessment)

VERSA maps intuitively depicted wavefront propagation, agreed broadly with activation maps, and illustrated conduction block, conduction breakthrough, and overall patterns of reentry

Original publication

VERSA: Omnipolar vector mapping from arbitrarily positioned electrodes. Heart Rhythm. Published 29 January 2026. doi:10.1016/j.hrthm.2026.01.032

Summarised from the published abstract; every number is checked against it