Authors' conclusion

For both CIED and ablation procedures the radiation with RSS was significantly lower than without RSS. Higher usage level brings higher reduction rates.

Study context

Population
Real-world electrophysiologic (EP) laboratory procedures (ablations and cardiovascular implantable electronic devices (CIED) procedures)35+19 without Radiaction Shielding System (RSS)
Exposure
RSS vs no RSS
Outcome
Radiation measured by sensors in different locations
Methodological detailsconsecutive real-life procedures
  • Consecutive real-life procedures

Main finding

As reported in the abstract

Radiation (procedures with ≥70% usage level)

Ablations: reduction in radiation with RSS

87%

76%–97% for different sensors

CIEDs: reduction in radiation with RSS

83%

59%–92%

Observational data: associations, not effects.

Further findings

RSS usage level

Average usage level, ablations
95%
Average usage level, CIEDs
88%

Procedure time and radiation time

RSS usage did not increase procedure time and radiation time

Original publication