Arrhythmias & electrophysiology

Autothreshold algorithm feasibility and safety in left bundle branch pacing

EuropacePublished Prospective comparative study

Authors' conclusion

The autothreshold algorithm is feasible in LBBP, with a favourable trend towards bipolar pacing. Automatic thresholds are similar to manual in patients with LBBP, and they remain stable during follow-up.

Study context

Population
Consecutive patients with indication for cardiac pacing
Participants
75 patients
Exposure
Left bundle branch pacing (LBBP) vs conventional pacing
  • LBBPn = 50
  • Conventional pacingn = 25reference
Outcome
Behaviour of the autothreshold algorithm (feasibility of activation; automatic vs manual thresholds)
Follow-up
Acute, subacute (1-7 days) and chronic (1-3 months) phases
Methodological detailsnon-randomized, single-centre
  • Non-randomized
  • Single-centre

Main finding

As reported in the abstract

LBBP: no statistically significant differences between manual and automatic thresholds; both stable during follow-up (mean increase)

0.50 V

Observational data: associations, not effects.

Further findings

Causes of failure of algorithm activation

Insufficient safety margins

Atrial tachyarrhythmias

Electrical noise LBBP 6.9%

Original publication

Autothreshold algorithm feasibility and safety in left bundle branch pacing. Europace. Published 3 December 2023. doi:10.1093/europace/euad359