Arrhythmias & electrophysiology

Cardiovascular implantable electronic device lead safety: Harnessing real-world remote monitoring data for medical device evaluation

Heart RhythmPublished Retrospective cohort study

Authors' conclusion

RM may strengthen real-world assessment of lead failure, particularly for leads where patients do not undergo replacement.

Study context

Population
Patients with a cardiovascular implantable electronic device under active remote monitoring, April 2019–April 2021
Participants
48,191 patients
Outcome
Incident lead abnormalities detected by RM and verified structural lead failure

Main finding

As reported in the abstract

Incident lead abnormalities detected by remote monitoring (RM)
1170 (2.4%)
RM abnormality with administrative codes: structural lead failure verified
233 of these 409 patients (57.0%)
RM abnormality without administrative codes: structural lead failure verified
167 (21.9%)
RM transmissions suggestive of lead abnormalities with structural lead failure
400 (34.2%)
Leads abandoned/extracted for structural failure without preceding abnormal RM transmissions
200
Total lead failure cohort (66.7% with RM abnormalities, 33.3% without)
600

Observational data: associations, not effects.

Further findings

Isolated right atrial or LV lead failure

Less likely to have lead replacement and administrative codes reflective of lead failure

Original publication

Cardiovascular implantable electronic device lead safety: Harnessing real-world remote monitoring data for medical device evaluation. Heart Rhythm. Published 28 December 2022. doi:10.1016/j.hrthm.2022.12.033