Electrocardiographic predictors of atrial mechanical sensing in leadless pacemakers

Heart RhythmPublished Prospective cohort study

Authors' conclusion

P-wave amplitudes in ECG leads aVL and V1 can predict A4 signal amplitude in patients with VDD LP and therefore the probability of successful AV synchronous pacing.

Study context

Population
Consecutive patients undergoing VDD leadless pacemaker implantation61 in A4 signal analysis
Participants
80 patients
Outcome
Association of preoperative 12-lead ECG P-wave amplitude in sinus rhythm with atrial mechanical sensing signal amplitude (A4)
Methodological details2-center, consecutive patients, 19 patients in VVI mode during follow-up excluded from A4 analysis
  • 2-center
  • Consecutive patients
  • 19 patients in VVI mode during follow-up excluded from A4 analysis

Main finding

As reported in the abstract

P-wave amplitude vs A4 signal amplitude
Absolute (aVL, aVF, V1, V2) and BMI-adjusted (I, II, aVL, aVF, aVR, V1, V2) P-wave amplitudes significantly positively correlated with A4 signal amplitude
all P <.05
Combined P-wave amplitude ≥0.2 mV in V1 and aVL predictive of A4 signal ≥1 m/s2: specificity
83% (95% CI 67%-100%)

Observational data: associations, not effects.

Further findings

Significant correlation of A4 signal amplitude and overall AV synchrony
P = 0.013

Original publication

Electrocardiographic predictors of atrial mechanical sensing in leadless pacemakers. Heart Rhythm. Published 1 March 2024. doi:10.1016/j.hrthm.2024.02.061

Composed by AI from the published abstract and checked against it. How Briefs are made →