Arrhythmias & electrophysiology

Impact of Device Compression on Peridevice Leak After Left Atrial Appendage Closure: The Impression Left Atrial Appendage Closure (LAAC) Study

JACC Clin ElectrophysiolPublished Retrospective cohort study

Authors' conclusion

In this dual-center cohort of TEE-guided LAAC, the overcompression group was independently associated to a higher rate of LAA occlusion at TEE follow-up.

Study context

Population
Consecutive patients undergoing transesophageal echocardiography (TEE)-guided LAAC with Watchman FLX in 2 high-volume centres, 2020–2023
Participants
61 vs 175
Exposure
Watchman FLX overcompression vs normal compression
  • Overcompressionn = 61
  • Normal compressionn = 175reference
Outcome
Residual peridevice leak (PDL) at TEE follow-up
Follow-up
2 months
Methodological detailsdual-centre, consecutive patients
  • Dual-centre
  • Consecutive patients

Main finding

As reported in the abstract

Residual PDL · 2-month TEE follow-up

P < 0.001

Observational data: associations, not effects.

Further findings

PDL at TEE follow-up, multivariate analysis (device compression rate)

Device compression rate the only variable independently associated with PDL risk (adjusted OR 0.935; 95% CI 0.927-0.948)
P = 0.003

CV death, ischemic stroke or systemic embolism · during follow-up

Not significantly different between groups (overcompression vs normal compression: 9.8% vs 12%)
P = 0.971

Safety

Procedural complications (composite)

Not significantly different between groups (overcompression vs normal compression: 1.6% vs 1.7%)
P = 0.968

Original publication

Impact of Device Compression on Peridevice Leak After Left Atrial Appendage Closure: The Impression LAAC Study. JACC Clin Electrophysiol. Published 2 October 2025. doi:10.1016/j.jacep.2025.08.014

Composed by AI from the published abstract and checked against it. How Briefs are made →

Get them every week

New studies on this topic in My Week and in your Monday e-mail.

We e-mail you a sign-in link and a code. No password, no spam.