Valvular & structural heart diseaseCohort study
Impact of Measured and Predicted Prosthesis-Patient Mismatch After Transcatheter Aortic Valve Replacement
- Registry cohort study
- multicentre
- international (26 centres)
Authors' conclusion
After adjusting for confounders, neither PPMm nor PPMp significantly affected 2-year all-cause mortality.
Study context
- Population
- Patients undergoing transcatheter aortic valve replacement (TAVR), IMPPACT TAVR registry, 26 international centres
- Participants
- 38,808 patients · 26 centres
- Exposure
- Measured vs predicted prosthesis-patient mismatch after TAVR
- Outcome
- All-cause mortality
- Follow-up
- 2 years
Main finding
As reported in the abstract
2-year all-cause mortality
Hazard ratio (95% CI)P value
Severe PPMm
1.22(1.02–1.45)
P = 0.027Severe PPMp
0.81(0.55–1.19)
P = 0.291Observational data: associations, not effects.
Further findings
Prevalence of PPM · prevalence significantly lower with PPMp
P < 0.001
Moderate
Severe
- Prevalence of PPM — Correlation between methods negligible (Kendall's tau c)
- 0.063
- Prevalence of PPM: Kendall's tau c P < 0.001
Original publication
Impact of Measured and Predicted Prosthesis-Patient Mismatch After Transcatheter Aortic Valve Replacement. JACC Cardiovasc Interv. Published 25 November 2024. doi:10.1016/j.jcin.2024.08.041
Summarised from the published abstract; every number is checked against it